How can I restore my old pH sensor that is slow, and gives a slope error (low span)?

/How can I restore my old pH sensor that is slow, and gives a slope error (low span)?

How can I restore my old pH sensor that is slow, and gives a slope error (low span)?

If the pH response becomes sluggish and the slope of the probe starts to get too low, IC Controls Electrode Renew Solution can be used (P/N A1100092). The electrode renew solution will skim off a micro-layer of the coated glass, thereby exposing a new glass surface which will exhibit better pH response.

2017-05-16T19:31:34+00:00 May 16th, 2017|