605 – SS Ball Valve Retractable pH Sensor

605 – SS Ball Valve Retractable pH Sensor

SS Body Ball Valve Retractable pH Sensor, with Lanyard Cables

The IC Controls model 605 was designed to provide reliable pH measurement in demanding applications where pressure cannot be reduced to zero for sensor removal.

Description

The model 605 meets the requirement for a ball valve insertion sensor in pulp & paper stock lines and other difficult applications.

The small cross-section steel housing with epoxy-encapsulated sensor better survives the rigors of direct insertion into dense stock. Larger-diameter sensors are more subject to pressures that can cause breakage during insertion and removal caused by bending of the sensor.

This sensor was originally requested as a replaceable cartridge style. However, cartridges experienced drifting and premature failure due to moisture reaching the BNC fitting on the back of the sensor cartridge. The Model 605 single-piece epoxy-encapsulated sensor eliminates the moisture problem and provides long, trouble-free operating life.

For Reliable Measurement Where Tough Processes Require

  • The strength of stainless steel at only 2 cm (¾ inch) diameter
  • Sealed, disposable design
  • Solution ground
  • Ball valve retraction to alleviate process interruption during calibration or service

Benefits

  • Sealed self-pressurizing design
  • Stainless steel body; optional titanium body and PVDF cartridge available
  • Stable annular reference chamber
  • Low maintenance
  • Easy calibration

Technical Data

Applications & Technical Articles

Download Article

pH/ORP Sensors in Metallurgical Applications

Critical Elements of the Sensor, Material Selection, Sensor Size Matters, Keeping Contaminants Out, Taking the Heat, Direct Connect versus Preamp.
Download Article
Download Article

pH Theory and Measurement

What is pH? Mathematical and descriptive definition, pH measurement techniques, pH sensing electrode, glass electrode, how glass electrodes work, types of glass electrodes, temperature affects on glass, reference electrode, referenced junction impact, ceramic junctions, wood junctions, etc.
Download Article

Customer Reviews

“I have test your pH electrode and that worked well!”

from a Canadian Consulting firm – Industrial Process Engineering experts in their field
using our Model 605 pH sensor in their Lithium Ion Battery Recycling application
May 2022

“Customer has been using IC Controls for years with a high level of satisfaction“

from our Distributor about their customer in the Pulp and Paper Industry
using our Model 655-8 pH analyzer and Model 605 pH sensor
May 2015

Application Review

Complete & Submit Questionnaire Now

Drawings

Wiring Diagram

pH Sensor to Analyzer Connection

Electrical Schematic for Model 602, 605, 606, 607, 614, 615, 617, 625, 627, 642, 643 and 648
Wiring Diagram
Assembly Drawing

Retractable ball valve sensor

A2100052 = option -21 for model 403 conductivity sensor and 605 pH sensor.
Assembly Drawing
For Illustration purposes only

Simple Bypass (side stream) setup

Showing isolation valve for safe sensor maintenance
For Illustration purposes only
Go to Top